Microprocessor Based Bench top Conductivity/TSD/SAL. MeterLIM-CON-01
May 15, 2016
Microprocessor Conductivity Pro Meter (EC-WALKLABARD)
May 15, 2016
Show all

Advance Microprocessor Conductivity Meter

Salient Features

  • LI-3200 EC Conductivity meter includes measurement of conductivity temperature in liquids

 

  • AUTOMATIC CALIBRATION

 

The meter is capable for calibrating of 1 point with using standard solution. From setup menu,

choose CAL (Calibration) and press key to set known conductivity value, immerse conductivity

electrode into calibration solution, press ENTER key and meter is automatically calibrated.

  • CONDUCTIVITY AUTO-RANGING

LI-3200 EC have auto range function for measuring conductivity. While immersing conductivity sensor

into sample solution, the screen immediately displays the value in appropriate measurement range for accuracy.

  • AUTOMATIC TEMPERATURE COMPENSATION

Temperature value is displayed on the screen. The meter automatically compensates for optimum accuracy

under variable temperature conditions.

  • ADJUSTABLE TEMPERATURE COEFFICIENT AND CELL CONSTANT

Various ions may have different temperature coefficients. The meters can set the temperature coefficient between 0 and 3.0% in degrees centigrade for optimum accuracy, set three cell constant (0.1, 1 and 10).

ACCESSORIES

  1. Conductivity Cell (K=1),
  2. Temperature Probe
  3. Electrode Stand
  4. 9V AC  adaptor
  5. Instruction manual

Specifications

Conductivity Range 0.00 to 200.0 ms (0.00 to 1.999, 19.99, 199.9, 1999 µS, 19.99) in five auto ranges
Conductivity Accuracy ±1% FS
Temperature Range 0 to 1000C
Temperature Accuracy ±10C
Temperature Coefficient 0 to 3.9% per 0C
Calibration Points Single Point
Cell Constant 0.1, 1.0, 10 (adjustable)
Temperature Compensation 0 to 500C, Manual or Automatic
Power Supply DC 9V adapters, using AC 220V/50 Hz
Dimensions 210 x 205 x 65
Weight 1.75 kg