Mortuary Chamber (LI-MOR-52)
May 24, 2016
MEDICO/CLINICAL CENTRIFUGE (LI-CEN-40)
May 24, 2016
Show all

Seed Germinator (Single/Dual Chamber)

SPECIFICATIONS

  • Reliable
  • Precise Control of Environmental Parameters
  • Micro-processor based Controllers
  • Aesthetically designed.
  • Corrosion Resistant interior and exteriors
  • Energy Efficient
  • CFC free
  • Sturdy Construction
  • Low Maintenance
  • Light Weight

CONSTRUCTION

Labard seed germinator are double walled convection heated and cooled units (Double/Single chamber Units). Outer body of our seed germinator are constructed out of thick PCRC sheet duly pre-treated with primers and rust proofing and painted with long lasting stove enamel or elegantly powder coated. The inner chamber is made of heavy gauge stainless steel sheet of SS-304 grade or anodized aluminum sheet or G.I. (as per the customer’s requirements). The gap between the walls is filled high grade mineral glass wool, which ensures maximum thermal efficiency in our walk in incubators.

The unit is provided with two doors, the inner door is made of thick flexi glass/float glass, to view the specimens/culture media/stocks, without disturbing the temperature of the chamber. This door is provided with magnetic door closer. The outer door is made of mild steel sheet lined with stainless steel from inside. This door is provided with lock and key arrangement. The unit is mounted on a sturdy steel frame The unit is provided with various customized shelves in various permutations and combinations to suit individual requirements. The triple walled back of our seed germinator are provided with two/Four/Six/Eight (size specific) air circulation fans for uniform maintenance of the temperature throughout the chamber. The unit is provided with 14 trays stainless steel (grades-304) in single chamber models and 28 stainless steel trays in double chambers. However we can customize the same as per the individual requirements.

 

HEATING

Indirect heating system is provided in our units, comprising of air heaters made of high grade Kanthal A-1 wires of suitable wattage. The warm air is evenly distributed throughout the chamber through efficient motor fans ensuring a very good temperature sensitivity.

 

Cooling  :

An energy efficient cooling unit is installed in our seed germinator to enable bio chemical demand, cyclic and growth studies at lower room temperatures. We use ISI marked high end CFC free compressors conforming to latest international standards and guidelines.

Humidity  :

Humidity generation provision is incorporated in our seed germinator The humidity is generated by means of aerosol humidity generator with efficient humidity controller cum indicator.

Humidity  Range :

From Ambient to 95% (As per the temperature requirements).

Humidity  Sensitivity :

Humidity is controlled by mean of an electronic humidity controller cum indicator with an accuracy of + 5%.

 

 

TEMPERATURE CONTROL

The temperature inside our seed germinator is controlled through programmable micro-processor based temperature controller cum indicator. Both the chambers have their individual control modules.

 

Temperature Range :

Temperature range of our standard seed germinator is 50 C to 600 C or 5o C above ambient to 600 C However we have the capacity to modify the same to suit the individual specialized requirements of our customers.

Temperature Sensitivity :

Temperatures inside our seed germinator are controlled with a sensitivity of + 1.00 C or better.

 

ILLUMINATION

Our seed germinator are provided with door operated illumination system comprising of fluorescent lights, beside two panels of diffused fluorescent light panel located at the either side of the chamber to enable the user to conduct the photo periodic tests and experiments.

 

Illumination Control :

The illumination is controlled through solid state cyclic timer to control the photoperiodic cycle of the chamber.

FORNT PANEL

Front panel of our units comprises of on/off switches heating, cooling and mains indicator lamps, temperature controllers and voltmeters.

TECHNICAL MATRIX

Safety Thermostat
Temperature Variation Adjustment 3 (With PID Controllers Only)
Temperature Sensors PT-100
Automatic Temperature Setting Yes (With PID Controllers Only)
Adjustable Limits Yes (With PID Controllers Only)
Light Control
Readability or Set Ability% 10% (optional feature)
Light Intensity (Middle Chamber) LUX As Desired (optional feature)
Light Intensity (Both Sides) LUX As Desired (optional feature)
Shelf
Standard Shelf 14 (Single Chamber)/28 Double Chamber
Internal Dimensions mm Model Specific
Maximum Permitted load per shelf kg 5 kg
Maximum Permitted total shelf kg 160 kg
Power Consumption
Nominal Power W 950
Nominal Voltage V 220-230Volts, 50Hz Single Phase
Frequency HZ 50

SPECIFICATIONS

Single / Double Chamber
Chamber of highly polished S.S. Capacity
555 (W) × 910 (H) × 605 (D) mm 308 liters
Heating and cooling in both chamber and chamber made of stainless steel sheet.

OPTIONAL

  • Automatic 0-24 hour’s timer to regulate cyclic illumination conditions.
  • Automatic voltage Stabilizer 3KVA (recommended).