Microprocessor Flame Photometers – LIM-204 are used for the determination of sodium, potassium, calcium and lithium. They use the latest microcontroller technology and advanced engineering techniques so as to give enhanced and reproducibility. They have soft touch membrane key for ease of operations.
The sample solution is aspirated through an atomizer. Sample, Air and the fuel are mixed in the mixing chamber which is then sprayed as a very fine mist into the flame. The color of the flame is changed depending upon the concentration of elements present. Radiations from the flame passes through the sensing system and specific narrow band interference filter which permits only the characteristic radiation to pass to the photo-detector. The output of the photo-detector is then processed by the microcontroller and the final results are displayed on the 20 x 4 line alphanumeric LCD with backlit.
In both models 5-Point calibration facility and non-linearity correction is provided through an automatic curve fitting software. Facility has been provided for re-standardization with a single standard, thereby, eliminating the need of the multiple standards once a calibration has been performed.
General | Serum | Urine | ||
Range | Na : 0.1-100 ppm Up to 250 meq/l, 1:100 dil
K : 0.1-100 ppm Up to 250 meq/l, 1:100 dil
Ca : 15-100 ppm Up to 250 meq/l, 1:100 dil
Li : 0.5-100 ppm Up to 250 meq/l, 1:100 dil
|
Na : 0-200 1:100 dil
K : 0-10 1:100 dil
Ca : – Li : 0-2 1:5 dil |
Na : 0-250 1:100 dil
K : 0-100 1:100 dil
Ca : – Li : 0-10 1:2 dil |
|
Sensitivity | General Mode : Na : 0.5ppm , K : 0.5ppm , Li : 0.5 ppm, Ca :15ppm | |||
Filters | Na & K Supplied (Ca & Li optional at extra cost) | |||
Resolution | 0.1 ppm / meq | |||
Reproducibility | ± 2% FS & ± 2 Digits | |||
Curve Fit Accuracy | ± 2 % FS | |||
Display | 20 x 4 Alphanumeric backlit LCD display | |||
Average Time | In-built in Software | |||
Flame System | LPG & dry oil free air | |||
Detector | Silicon Photodiode | |||
Calibration | Up to 5-Point Calibration with curve fitting software | |||
Gas Control | Adjustable with knobs | |||
Ignition System | Auto Ignition | |||
Atomizer | Axial flow type | |||
RS-232 Interface | Available (PC Software at extra cost) | |||
Power | 230 V, ±10 % AC, 50 Hz |